EQU-3937 - USA (Columbia, South Carolina) - Gas Chromatograph, Mass Spectrophotometer and Data System - Deadline May 6,2020

$
0
0
Posted Date : April 24,2020
Product (RFP/RFQ/RFI/Solicitation/Tender/Bid Etc.) ID : EQU-3937
Government Authority located in Columbia, South Carolina; USA based organization looking for expert vendor for gas chromatograph, mass spectrophotometer and data system.
[A] Budget: looking For Proposal
[B] Scope of Service:
Vendor needs to provide gas chromatograph, mass spectrophotometer and data system
- Gas Chromatograph must have the following minimum specifications and include a column oven, injection port and autosampler:
- Must have capability to accurately and reproducibly execute up to 20 oven ramps and 21 oven plateaus in a run,
- Must be able to execute negative temperature ramps,
- Must be able to execute three pressure / flow ramps in a run,
- Functional keyboard with a four-line alphanumeric display.
- Run time programming, Clock time programming, Stopwatch function,
- System must automatically shut off gases in event of a leak,
- Must have built in, permanently calibrated barometer and thermometer to compensate for ambient laboratory changes for maximum reproducibility,
- Mass Spectrophotometer - The system must have a Mass Selective Detector included which meets the following minimum specifications:
- injected in 1uL iso-octane solvent onto a 30m x 0.25mm x 0.25um HP-5MS ultra inert column with data acquired at 2.9 scans/second over the mass range 50-300 amu shall yield an RMS signal/noise of at least 550:1 for the extracted ion signal at mass 272.0, using helium as the carrier gas and a diffusion vacuum system,
- The mass spectrophotometer must use an ion source where the metallic parts are constructed from inert material. Ion source must be controllable from 100-350 degrees C. Stainless steel is NOT acceptable nor is coating the metallic parts with an inert material,
- The data system must be able to automatically upload compatible data files to existing technologies data processing workstations for data review and production of finished reports, The software must have computer-based training for self- paced training,
- The software is to allow complete control of all operating parameters; including control of all the pressure programming capabilitisssses of the electronic pneumatics control capillary injection port,
- Data analysis should allow individual spectra to be selected, subtracted, and searched against either published or user created libraries,
- System must have the capability for the user to add additional libraries or to create custom libraries from their own spectra, etc. to the government authority located in Columbia, SC.
[C] Eligibility:
Onshore (USA Only);
[D] Work Performance:
Performance of the work will be Offsite. Vendor needs to carry work in their office location.
Budget :
Deadline to Submit Proposals: May 06,2020
Cost to Download This RFP/RFQ/RFI/Solicitation/Tender/Bid Document : 5 US$